Seimens 1.5 Vision Scanner
Sequence file name: fi3d_tun_6rb195.xkc
TR 36
TE 6
FLIP ANGLE 20 DEGREES
# OF SLABS 1
SLAB THICKNESS 60
EFF. THICKNESS .86
DISTANCE FACTOR 0
ORIENTATION TRANSVERSE
# OF PARTITIONS 70
MATRIX 256X256
REC FOV 6/8
FOV 220
# OF ACQ. 1
# OF MEASURES 1
PHASE DIRECTION R/L

GE 1.5 Excite Twin Speed Scanner
Pulse Sequence FSPGR
IMAGING OPTIONS - VBW, FAST, IRPREP, ZIP2
TR NOT SELECTABLE
TE MINIMUM
PREP TIME 450
FLIP ANGLE 20 DEGREES
PLANE TRANSVERSE
# OF SLABS 1
SLAB LOCATIONS 80
SLICE THICKNESS .8
PHASE/FREQUENCY 224/320
FOV 20
PHASE FOV .9
NEX 1
FREQ. DIRECTION UNSWAPPED
BAND WIDTH 16.67
GRADIENT MODE ZOOM
MODE 3D

Position patient in head coil as for a routine Brain exam.
Inject 15 ml of Gadolinium I.V.
Acquire a standard 3 plane Brain scout sequence.
Position Trigeminal Sequence with the bottom of the Slab at the base of the Skull.
Be sure to adjust for head tilt. This should give you adequate coverage to see the
5 th Cranial Nerve.
Post Processing can be done in the MPR (multi planar reconstruction) mode
specific to the scanner being used. Reconstructing the transverse images in
sagittal and coronal planes, will help in evaluating nerve and vessel placement.